== Amptek Si-pin ==
== Amptek Si-pin ==


This data is just for reference, and here is a picture of the device with bond \ wires attached to a pcb for testing purposes. [[image:ampteksipin.jpg]]
This data is just for reference, and here is a picture of the device with bond \
wires attached to a pcb for testing purposes.
[[file:ampteksipin.jpeg|thumb|border|Closeup mounted]]
 
Next I collected V-I data on this detector:
 
Data acquired on 6/10/2014 using bench supplies at room temp:
{| align="center boarder = "1"
! Voltage (no load, V)
! Voltage (diode load, V)
! Current (measured, uA)
|-
! Reverse bias
|-
| 10 || 5.7
|-
| 20 || 5.8
|-
| 30 || 5.9
|-
| 40 || 6.0
|-
| 50 || 6.0
|-
| 60 || 6.1
|-
| 70 || 6.1
|-
| 80 || 6.2
|-
| 90 || 6.2
|-
| 100 || 6.3
|-
| 110 || 6.3
|-
| 120 || 6.3
|-
| 130 || 6.4
|-
| 140 || 6.4
|-
| 150 || 6.4
|-
| 160 || 6.4
|-
| 170 || 6.5
|-
| 180 || 6.5
|-
| 190 || 6.5
|-
| 200 || 6.5
|-
| 210 || 6.6
|-
| 220 || 6.6
|-
| 230 || 6.6
|-
| 240 || 6.7
|-
| 250 || 6.7
|-
| 260 || 6.7
|-
| 270 || 6.7
|-
| 280 || 6.8
|-
| 290 || 6.8
|-
| 300 || 6.9
|}
 
Data acquired on 6/10/2014 using bench supplies at room temp:
{| align="center boarder = "1"
! Voltage (no load, V)
! Voltage (diode load, V)
! Current (measured, uA)
|-
! Reverse bias
|-
| 0.1 || 4.9
|-
| 0.2 || 4.0
|-
| 0.3 || 3.8
|-
| 0.4 || 23.6
|-
| 0.5 || 35.4
|-
| 0.6 || 60.0
|-
| 0.7 || 73.0
|-
| 0.8 || 102
|-
| 0.9 || 112
|-
| 1.0 || 127
|-
| 1.1 || 153
|-
| 1.2 || 181
|}

Latest revision as of 16:23, 21 March 2023

Amptek Si-pin[edit | edit source]

This data is just for reference, and here is a picture of the device with bond \ wires attached to a pcb for testing purposes.

Closeup mounted

Next I collected V-I data on this detector:

Data acquired on 6/10/2014 using bench supplies at room temp:

Voltage (no load, V) Voltage (diode load, V) Current (measured, uA)
Reverse bias
10 5.7
20 5.8
30 5.9
40 6.0
50 6.0
60 6.1
70 6.1
80 6.2
90 6.2
100 6.3
110 6.3
120 6.3
130 6.4
140 6.4
150 6.4
160 6.4
170 6.5
180 6.5
190 6.5
200 6.5
210 6.6
220 6.6
230 6.6
240 6.7
250 6.7
260 6.7
270 6.7
280 6.8
290 6.8
300 6.9

Data acquired on 6/10/2014 using bench supplies at room temp:

Voltage (no load, V) Voltage (diode load, V) Current (measured, uA)
Reverse bias
0.1 4.9
0.2 4.0
0.3 3.8
0.4 23.6
0.5 35.4
0.6 60.0
0.7 73.0
0.8 102
0.9 112
1.0 127
1.1 153
1.2 181